Cable break and damaged strain relief.
Depending on the model, the probe cable carries more than a hundred fine wires, one per element. At the strain relief — where the cable leaves the grip — and at the connector, the cable is bent during every examination. If individual wires break, crosstalk arises between the channels.
How to spot it
The image flickers or shows artefacts that look like element failure but change position when someone moves the cable. That is the test: carefully bend the cable at the strain relief and watch the reverberation image. The housing is usually completely intact.
What we do
Sections of the cable can be renewed, as can the strain relief and the connector. If the break affects the element leads in the bundle, it becomes a more extensive repair. Which route is needed is shown by measurement — not by estimate.
Usual repair level
Every repair is individual and quoted on request — depending on the probe model and its condition. After the analysis you receive a transparent quotation. Additional services: disinfection to ISO standards, collection across Europe, 24h express service.
Probe types affected
- Standard & matrix probes
- 3D/4D & matrix 3D/4D
- TEE & matrix TEE probes
Send in a probe with this fault
Five questions in the repair advisor, and at the end you get a completed enquiry form with an initial assessment.
Request a repairWhich manufacturers
This fault pattern occurs across all manufacturers. To the probes of yours: