Delaminated lens.
The acoustic lens is bonded to the transducer. If that bond comes apart, an air gap forms — and air reflects ultrasound almost completely. Triggers are disinfectants, mechanical strain and simply age. In quality-assurance studies delamination is by far the most common single finding.
How to spot it
The image goes flat, penetration depth drops, contrast and detail resolution suffer — without anything looking broken from the outside. Often it is only noticed when someone performs the same examination with a second probe. Under raking light, blisters or waves in the lens can sometimes be seen.
What we do
We remove the old coating and build it up again. On volume probes and endocavity probes refurbishing the housing is part of it. If the transducer underneath is already damaged, we say so before the repair instead of putting a coating over a dead element.
Usual repair level
Every repair is individual and quoted on request — depending on the probe model and its condition. After the analysis you receive a transparent quotation. Additional services: disinfection to ISO standards, collection across Europe, 24h express service.
Probe types affected
- Standard & matrix probes
- Wireless probes
- 3D/4D & matrix 3D/4D