Image dropout and shadowing.
Technically this is element failure, in English dropout. Individual transducer elements no longer transmit or receive — because an element lead has broken, because the crystal has cracked after a drop, or because the lens above the element has delaminated. In the image a wedge-shaped gap appears that widens with depth.
How to spot it
Hold the probe freely in the air and look at the reverberation image: a vertical dark stripe that stays in the same place when you move the probe is element failure. Also change the port on the device — if the stripe stays, it is not the connection. Switch off harmonic imaging and speckle reduction, both mask exactly this finding.
What we do
We measure every element individually for sensitivity and capacitance and so narrow down whether the lead, the crystal or the coating is affected. Depending on the finding, repairing the element leads is enough, or it becomes a full overhaul with crystal replacement. The result is in the quotation before you decide.
Usual repair level
Every repair is individual and quoted on request — depending on the probe model and its condition. After the analysis you receive a transparent quotation. Additional services: disinfection to ISO standards, collection across Europe, 24h express service.
Probe types affected
- Standard & matrix probes
- Wireless probes
- 3D/4D & matrix 3D/4D
- TEE & matrix TEE probes
- Ultrasound endoscopes
Send in a probe with this fault
Five questions in the repair advisor, and at the end you get a completed enquiry form with an initial assessment.
Request a repairWhich manufacturers
This fault pattern occurs across all manufacturers. To the probes of yours: